Q&A
about VFS-D7 Fusion Splicer
1>The
screen is dark after starting machine and the level lamp on electrodes
doesn¡¯t work after opening windshield.
Solutions:
1)
Check whether the electrical outlet is inserted.
2)
Check whether the fuse is broken, and then replace the broken one.
2> ¡°Exhausted Battery¡± is
displayed on the bottom of screen after starting
VFS-D7 and buzzer
works continually.
Solutions:
If use battery, then
replace the power supply.
3> The fiber can be reset as
normal, but when setting gap, fiber is immovable
on the screen and
¡°Replace Fiber¡± is displayed on the screen.
Solutions:
1) Press ¡°¡±
to reset system.
2) Raise the windshield,
left and right clamps. The checking steps as follows:
3) Check whether broken
fiber exists.
4) Check whether cleaved
length is too short.
5) Check whether fiber
matches with V-groove.
4> The fiber can be reset
as normal, but when setting gap, fiber moves
backwards
continually, ¡°Replace fiber¡± are displayed on screen.
Solutions:
Maybe there is too much
dust on the lens of microscope, clean the aclinic
and upright lens with
cotton swab until dust disappears, then try again.
5> The fiber can be reset
as normal, but when setting gap, ¡°Replace fiber¡± is displayed on
screen.
Solutions:
1) Press ¡°¡±
to reset system.
2) Raise the windshield,
left and right clamps. The checking steps as follows:
3) Check whether broken
fiber exists.
4) Check whether cleaved
length is too short.
5) Check whether fiber
matches with V-groove
6> When calibrating
automatically, fiber moves up and down, the screen¡¯s
display stops at ¡°Adjust core¡±.
Solutions:
1) Press ¡°¡±
to reset system.
2) Check whether bare
fiber is clean.
3) Clean the dust of
V-groove.
4) Knock at the holder
until fiber is impacted, then try again.
7>The fiber can be reset as
normal, but when setting gap, ¡°Bad left fiber¡¯s surface¡± is
displayed.
Solutions:
1)
Observe fiber picture in screen, if the quality of left fiber¡¯s
surface is bad£¬you
can remake surface and try again.
2)
Observe fiber picture in screen, if the quality of left fiber¡¯s
surface is good,
maybe the value of ¡°Setting surface¡± is too small, you can increase
it.
3)
After ¡°Bad left fiber¡¯s surface¡± is displayed, the screen becomes
dark;
Make sure
that windshield is covered well.
Raise
windshield, check whether the reeds are distorting, and then manage
it.
8> When
splicing the MM fiber, there is bubble in the process of arc.
Solutions: The difference between fiber¡¯s core and envelope¡¯s
refractive index
is too big. Operations as follows:
1)Consult
MM fiber¡¯s arc program set by factory and confirm.
2)Do arc
test until ¡°Moderate current¡± is displayed three times.
3)Splice
MM fiber; if there is bubble then modify arc parameter.
Operations are as follows:
1)
Enter the menu of arc parameter.
2)
Increase the value of ¡°Pre-fusion time¡± until no bubble.
3)
Decrease the value of ¡°Pre-fusion current¡± if no bubble and fiber
thins. |